The biocompatibility of nickel titanium (NiTi) wires joined to stainless steel (SS) wires via laser autogenous brazing has been evaluated. The laser joining process is designed to limit the amount of mixing of the materials, thus preventing the formation of brittle intermetallic phases. This process has the potential for manufacturing implantable medical devices; therefore, the biocompatibility must be determined. Laser joined samples underwent nickel release rate, polarization, hemolysis, and cytotoxicity testing. Competing effects regarding grain refinement and galvanic effects were found to influence the corrosion response. After 15 days of exposure to a simulated body fluid, the total nickel released is less than 2 ug/cm2. Numerical modeling of the corrosion currents along the wires, by making use of polarization data, helped to explain these results. Microbiological testing found a maximum hemolytic index of 1.8, while cytotoxicity tests found a zero toxicity grade. All of these results indicate that the autogenous laser brazing process results in joints with good biocompatibility.
Skip Nav Destination
Article navigation
June 2015
Research-Article
Biocompatibility and Corrosion Response of Laser Joined NiTi to Stainless Steel Wires
Grant Brandal,
Grant Brandal
1
Mechanical Engineering Department,
New York, NY 10023
e-mail: gbb2114@columbia.edu
Columbia University
,500 W 120th Street, Mudd Rm 220
,New York, NY 10023
e-mail: gbb2114@columbia.edu
1Corresponding author.
Search for other works by this author on:
Syed Naveed
Syed Naveed
Search for other works by this author on:
Grant Brandal
Mechanical Engineering Department,
New York, NY 10023
e-mail: gbb2114@columbia.edu
Columbia University
,500 W 120th Street, Mudd Rm 220
,New York, NY 10023
e-mail: gbb2114@columbia.edu
Y. Lawrence Yao
Syed Naveed
1Corresponding author.
Contributed by the Manufacturing Engineering Division of ASME for publication in the JOURNAL OF MANUFACTURING SCIENCE AND ENGINEERING. Manuscript received May 6, 2014; final manuscript received January 20, 2015; published online March 5, 2015. Assoc. Editor: Wei Li.
J. Manuf. Sci. Eng. Jun 2015, 137(3): 031015 (9 pages)
Published Online: June 1, 2015
Article history
Received:
May 6, 2014
Revision Received:
January 20, 2015
Online:
March 5, 2015
Citation
Brandal, G., Lawrence Yao, Y., and Naveed, S. (June 1, 2015). "Biocompatibility and Corrosion Response of Laser Joined NiTi to Stainless Steel Wires." ASME. J. Manuf. Sci. Eng. June 2015; 137(3): 031015. https://doi.org/10.1115/1.4029766
Download citation file:
Get Email Alerts
Cited By
Related Articles
Laser Autogenous Brazing of Biocompatible, Dissimilar Metals in Tubular Geometries
J. Manuf. Sci. Eng (April,2017)
Performance Study of Nickel Covered by Lithium Cobaltite Cathode for Molten Carbonate Fuel Cells: A Comparison in Li/K and Li/Na Carbonate Melts
J. Fuel Cell Sci. Technol (April,2010)
Chemical Treatment Method for the Aluminum Bipolar Plates of PEM Fuel Cells
J. Fuel Cell Sci. Technol (August,2005)
Strength and Phase Identification of Autogenous Laser Brazed Dissimilar Metal Microjoints
J. Manuf. Sci. Eng (February,2015)
Related Proceedings Papers
Related Chapters
Mathematical Background
Vibrations of Linear Piezostructures
Effects of Metallic Plate and Objects on Performance of Inverted F Antenna for ISM Band Application
International Conference on Computer and Automation Engineering, 4th (ICCAE 2012)
The Study on Linear Component Technique in Identifying the Polarization of Electromagnetic Waves
International Conference on Computer and Electrical Engineering 4th (ICCEE 2011)