Silicone implants are used for prosthetic arthroplasty of metacarpophalangeal (MCP) joints severely damaged by rheumatoid arthritis. Different silicone elastomer MCP implant designs have been developed, including the Swanson and the NeuFlex implants. The goal of this study was to compare the in vitro mechanical behavior of Swanson and NeuFlex MCP joint implants. Three-dimensional (3D) finite element (FE) models of the silicone implants were modeled using the commercial software ANSYS and subjected to angular displacement from 0 deg to 90 deg. FE models were validated using mechanical tests of implants incrementally bent from 0 deg to 90 deg in a joint simulator. Swanson size 2 and 4 implants were compared with NeuFlex size 10 and 30 implants, respectively. Good agreement was observed throughout the range of motion for the flexion bending moment derived from 3D FE models and mechanical tests. From 30 deg to 90 deg, the Swanson 2 demonstrated a greater resistance to deformation than the NeuFlex 10 and required a greater bending moment for joint flexion. For larger implant sizes, the NeuFlex 30 had a steeper moment-displacement curve, but required a lower moment than the Swanson 4, due to implant preflexion. On average, the stress generated at the implant hinge from 30 deg to 90 deg was lower in the NeuFlex than in the Swanson. On average, starting from the neutral position of 30 deg for the preflexed NeuFlex implant, higher moments were required to extend the NeuFlex implants to 0 deg compared with the Swanson implants, which returned spontaneously to resting position. Implant toggling within the medullary canals was less in the NeuFlex than in the Swanson. The differential performance of these implants may be useful in implant selection based on the preoperative condition(s) of the joint and specific patient functional needs.
Skip Nav Destination
e-mail: amirah@bu.edu
e-mail: jcstranart@hotmail.com
e-mail: meguid@mie.utoronto.ca
e-mail: bogoche@smh.ca
Article navigation
February 2011
Technical Briefs
Biomechanical Validation of Finite Element Models for Two Silicone Metacarpophalangeal Joint Implants
A. I. Hussein,
A. I. Hussein
Department of Mechanical and Industrial Engineering, Engineering Mechanics and Design Laboratory,
e-mail: amirah@bu.edu
University of Toronto
, Toronto, ON, M5S 3G8, Canada; Department of Mechanical Engineering, Boston University
, 110 Cummington Street, Boston, MA 02215
Search for other works by this author on:
J. C. Stranart,
J. C. Stranart
Department of Mechanical and Industrial Engineering, Engineering Mechanics and Design Laboratory,
e-mail: jcstranart@hotmail.com
University of Toronto
, Toronto, ON, M5S 3G8, Canada; University of Toronto
, 63 Norgrove Crescent, Toronto, ON, M9P 3C7, Canada
Search for other works by this author on:
S. A. Meguid,
S. A. Meguid
Professor
Department of Mechanical and Industrial Engineering, Engineering Mechanics and Design Laboratory,
e-mail: meguid@mie.utoronto.ca
University of Toronto
, Toronto, ON, M5S 3G8, Canada; Department of Mechanical and Industrial Engineering, University of Toronto
, 5 King’s College Road, Toronto, ON, M5S 3G8, Canada
Search for other works by this author on:
E. R. Bogoch
E. R. Bogoch
Department of Surgery,
e-mail: bogoche@smh.ca
University of Toronto
, Toronto, ON, M5C 1R6, Canada; Li Ka Shing Knowledge Institute and the Mobility Program Clinical Research Unit, St. Michael’s Hospital
, Toronto, ON, M5C 1R6, Canada
Search for other works by this author on:
A. I. Hussein
Department of Mechanical and Industrial Engineering, Engineering Mechanics and Design Laboratory,
University of Toronto
, Toronto, ON, M5S 3G8, Canada; Department of Mechanical Engineering, Boston University
, 110 Cummington Street, Boston, MA 02215e-mail: amirah@bu.edu
J. C. Stranart
Department of Mechanical and Industrial Engineering, Engineering Mechanics and Design Laboratory,
University of Toronto
, Toronto, ON, M5S 3G8, Canada; University of Toronto
, 63 Norgrove Crescent, Toronto, ON, M9P 3C7, Canadae-mail: jcstranart@hotmail.com
S. A. Meguid
Professor
Department of Mechanical and Industrial Engineering, Engineering Mechanics and Design Laboratory,
University of Toronto
, Toronto, ON, M5S 3G8, Canada; Department of Mechanical and Industrial Engineering, University of Toronto
, 5 King’s College Road, Toronto, ON, M5S 3G8, Canadae-mail: meguid@mie.utoronto.ca
E. R. Bogoch
Department of Surgery,
University of Toronto
, Toronto, ON, M5C 1R6, Canada; Li Ka Shing Knowledge Institute and the Mobility Program Clinical Research Unit, St. Michael’s Hospital
, Toronto, ON, M5C 1R6, Canadae-mail: bogoche@smh.ca
J Biomech Eng. Feb 2011, 133(2): 024501 (6 pages)
Published Online: January 31, 2011
Article history
Received:
February 3, 2010
Revised:
December 19, 2010
Posted:
December 22, 2010
Published:
January 31, 2011
Online:
January 31, 2011
Citation
Hussein, A. I., Stranart, J. C., Meguid, S. A., and Bogoch, E. R. (January 31, 2011). "Biomechanical Validation of Finite Element Models for Two Silicone Metacarpophalangeal Joint Implants." ASME. J Biomech Eng. February 2011; 133(2): 024501. https://doi.org/10.1115/1.4003311
Download citation file:
Get Email Alerts
Cited By
Improvement in Active Cell Proliferation Area at Higher Permeability With Novel TPMS Lattice Structure
J Biomech Eng (November 2024)
Modeling Fatigue Failure of Cartilage and Fibrous Biological Tissues Using Constrained Reactive Mixture Theory
J Biomech Eng (December 2024)
A Numerical Study of Crack Penetration and Deflection at the Interface Between Peritubular and Intertubular Dentin
J Biomech Eng (December 2024)
Related Articles
Design of a Dynamic Stabilization Spine Implant
J. Med. Devices (June,2009)
A Design Framework of Unloaded Leaflet Shape for the Ovine Pulmonary Valve Single Leaflet Replacement Surgery
J. Med. Devices (June,2011)
A New External Fixator Design for Femoral Fracture Reduction
J. Med. Devices (June,2008)
Finite Element Modeling of Resurfacing Hip Prosthesis: Estimation of Accuracy Through Experimental Validation
J Biomech Eng (February,2010)
Related Proceedings Papers
Related Chapters
Data Tabulations
Structural Shear Joints: Analyses, Properties and Design for Repeat Loading
Vibration Analysis of the Seated Human Body in Vertical Direction
International Conference on Computer Technology and Development, 3rd (ICCTD 2011)
Subsection NB—Class 1 Components
Companion Guide to the ASME Boiler & Pressure Vessel Codes, Volume 1 Sixth Edition