A custom program for the processing of pressure sensitive (Fuji) film data is presented and validated in this paper. Some of the shortcomings of previous descriptions of similar programs in literature are addressed. These shortcomings include incomplete descriptions of scan resolution, processing technique, and accuracy of results. Of these, the accuracy of results is the most important and is addressed in this study by using Fuji film calibration data. In Fuji film calibration, known loads are applied to forms with known area. The accuracy of this program and that of the two commercially available image processing programs were determined. The results of the custom program are found to be within 10% of the results from the commercial programs and from experimental data. This level of accuracy is the same reported level of accuracy of Fuji film, verifying the custom program for use in Fuji film contact pressure and area measurements.
Skip Nav Destination
e-mail: mgmst20@pitt.edu
Article navigation
January 2009
Technical Briefs
Characterization of the Performance of a Custom Program for Image Processing of Pressure Sensitive Film
Muturi G. Muriuki, Ph.D.,
Muturi G. Muriuki, Ph.D.
Department of Orthopaedic Surgery,
e-mail: mgmst20@pitt.edu
University of Pittsburgh
, Pittsburgh, PA 15261
Search for other works by this author on:
Lars G. Gilbertson, Ph.D.,
Lars G. Gilbertson, Ph.D.
Department of Orthopaedic Surgery,
University of Pittsburgh
, Pittsburgh, PA 15261
Search for other works by this author on:
Christopher D. Harner, M.D.
Christopher D. Harner, M.D.
Department of Orthopaedic Surgery,
University of Pittsburgh
, Pittsburgh, PA 15261
Search for other works by this author on:
Muturi G. Muriuki, Ph.D.
Department of Orthopaedic Surgery,
University of Pittsburgh
, Pittsburgh, PA 15261e-mail: mgmst20@pitt.edu
Lars G. Gilbertson, Ph.D.
Department of Orthopaedic Surgery,
University of Pittsburgh
, Pittsburgh, PA 15261
Christopher D. Harner, M.D.
Department of Orthopaedic Surgery,
University of Pittsburgh
, Pittsburgh, PA 15261J Biomech Eng. Jan 2009, 131(1): 014503 (4 pages)
Published Online: November 21, 2008
Article history
Received:
August 28, 2007
Revised:
August 4, 2008
Published:
November 21, 2008
Citation
Muriuki, M. G., Gilbertson, L. G., and Harner, C. D. (November 21, 2008). "Characterization of the Performance of a Custom Program for Image Processing of Pressure Sensitive Film." ASME. J Biomech Eng. January 2009; 131(1): 014503. https://doi.org/10.1115/1.3005150
Download citation file:
Get Email Alerts
How Irregular Geometry and Flow Waveform Affect Pulsating Arterial Mass Transfer
J Biomech Eng (December 2024)
Phenomenological Muscle Constitutive Model With Actin–Titin Binding for Simulating Active Stretching
J Biomech Eng (January 2025)
Image-Based Estimation of Left Ventricular Myocardial Stiffness
J Biomech Eng (January 2025)
Related Articles
Total Strain Fields of the Antero-Inferior Shoulder Capsule Under Subluxation: A Stereoradiogrammetric Study
J Biomech Eng (October,2001)
Three Dimensional Imaging of LIGA-Made Microcomponents
J. Manuf. Sci. Eng (November,2004)
Biomechanical and Elastographic Analysis of Mesenchymal Stromal Cell Treated Tissue Following Surgery
J Biomech Eng (July,2010)
Design of a Calibration Phantom for Measuring the Temporal Resolution of a Tomographic Imaging Device
J. Med. Devices (September,2007)
Related Proceedings Papers
Related Chapters
Dimensional Measurements and Calibration
Metrology and Instrumentation: Practical Applications for Engineering and Manufacturing
An Algorithm Improving Frequency Resolution Based on DDS Calibration
International Conference on Information Technology and Computer Science, 3rd (ITCS 2011)
Mechanical Measurements and Calibration
Metrology and Instrumentation: Practical Applications for Engineering and Manufacturing