Design of a Wireless Medical Capsule for Measuring the Contact Pressure Between a Capsule and the Small Intestine

[+] Author and Article Information
Pengbo Li

Mechanical and Material Engineering,
University of Nebraska-Lincoln,
W342 Nebraska Hall,
Lincoln, NE 68588-0526
e-mail: pengbo.li@huskers.unl.edu

Craig Kreikemeier-Bower

School of Veterinary,
University of Nebraska-Lincoln,
Mussehl Hall 110,
Lincoln, NE 68583

Wanchuan Xie

Mechanical and Material Engineering,
University of Nebraska-Lincoln,
W342 Nebraska Hall,
Lincoln, NE 68588-0526

Vishal Kothari

Department of Surgery,
University of Nebraska Medical Center,
Nebraska Medicine—Bariatrics Center
4400 Emile Street,
Omaha, NE 68198

Benjamin S. Terry

Department of Mechanical and
Materials Engineering,
University of Nebraska-Lincoln,
360 Walter Scott Engineering Center,
Lincoln, NE 68508

Manuscript received November 2, 2016; final manuscript received March 4, 2017; published online March 24, 2017. Assoc. Editor: Nathan Sniadecki.

J Biomech Eng 139(5), 051003 (Mar 24, 2017) (8 pages) Paper No: BIO-16-1433; doi: 10.1115/1.4036260 History: Received November 02, 2016; Revised March 04, 2017

A wireless medical capsule for measuring the contact pressure between a mobile capsule and the small intestine lumen was developed. Two pressure sensors were used to measure and differentiate the contact pressure and the small intestine intraluminal pressure. After in vitro tests of the capsule, it was surgically placed and tested in the proximal small intestine of a pig model. The capsule successfully gathered and transmitted the pressure data to a receiver outside the body. The measured pressure signals in the animal test were analyzed in the time and frequency domains, and a mathematic model was presented to describe the different factors influencing the contact pressure. A novel signal processing method was applied to isolate the contraction information from the contact pressure. The result shows that the measured contact pressure was 1.08 ± 0.08 kPa, and the small intestine contraction pressure's amplitude and rate were 0.29 ± 0.046 kPa and 12 min−1. Moreover, the amplitudes and rates of pressure from respiration and heartbeat were also estimated. The successful preliminary evaluation of this capsule implies that it could be used in further systematic investigation of small intestine contact pressure on a mobile capsule-shaped bolus.

FIGURES IN THIS ARTICLE
<>
Copyright © 2017 by ASME
Topics: Pressure , Design
Your Session has timed out. Please sign back in to continue.

References

Moglia, A. , Menciassi, A. , Schurr, M. O. , and Dario, P. , 2006, “ Wireless Capsule Endoscopy: From Diagnostic Devices to Multipurpose Robotic Systems,” Biomed. Microdevices, 9(2), pp. 235–243. [CrossRef]
Ciuti, G. , Menciassi, A. , and Dario, P. , 2011, “ Capsule Endoscopy: From Current Achievements to Open Challenges,” IEEE Rev. Biomed. Eng., 4, pp. 59–72. [CrossRef] [PubMed]
McCaffrey, C. , Chevalerias, O. , O'Mathuna, C. , and Twomey, K. , 2008, “ Swallowable-Capsule Technology,” IEEE Pervasive Comput., 7(1), pp. 23–29. [CrossRef]
Woods, S. P. , and Constandinou, T. G. , 2013, “ Wireless Capsule Endoscope for Targeted Drug Delivery: Mechanics and Design Considerations,” IEEE Trans. Biomed. Eng., 60(4), pp. 945–953. [CrossRef] [PubMed]
Xie, W. , Kothari, V. , and Terry, B. S. , 2015, “ A Bio-Inspired Attachment Mechanism for Long-Term Adhesion to the Small Intestine,” Biomed. Microdevices, 17(4), p. 68. [CrossRef] [PubMed]
Yim, S. , Gultepe, E. , Gracias, D. H. , and Sitti, M. , 2014, “ Biopsy Using a Magnetic Capsule Endoscope Carrying, Releasing, and Retrieving Untethered Microgrippers,” IEEE Trans. Biomed. Eng., 61(2), pp. 513–521. [CrossRef] [PubMed]
Slawinski, P. R. , Obstein, K. L. , and Valdastri, P. , 2015, “ Emerging Issues and Future Developments in Capsule Endoscopy,” Tech. Gastrointest. Endoscopy, 17(1), pp. 40–46. [CrossRef]
Yim, S. , and Sitti, M. , 2012, “ Design and Rolling Locomotion of a Magnetically Actuated Soft Capsule Endoscope,” IEEE Trans. Rob., 28(1), pp. 183–194. [CrossRef]
Kim, J.-Y. , Kwon, Y.-C. , and Hong, Y.-S. , 2012, “ Automated Alignment of Rotating Magnetic Field for Inducing a Continuous Spiral Motion on a Capsule Endoscope With a Twistable Thread Mechanism,” Int. J. Precis. Eng. Manuf., 13(3), pp. 371–377. [CrossRef]
Terry, B. S. , Schoen, J. A. , and Rentschler, M. E. , 2012, “ Characterization and Experimental Results of a Novel Sensor for Measuring the Contact Force From Myenteric Contractions,” IEEE Trans. Biomed. Eng., 59(7), pp. 1971–1977. [CrossRef] [PubMed]
Francisco, M. M. , Terry, B. S. , Schoen, J. A. , and Rentschler, M. E. , 2016, “ Intestinal Manometry Force Sensor for Robotic Capsule Endoscopy: An Acute, Multi-Patient In Vivo Animal and Human Study,” IEEE Trans. Biomed. Eng., 63(5), pp. 943–951. [CrossRef] [PubMed]
Terry, B. S. , Lyle, A. B. , Schoen, J. A. , and Rentschler, M. E. , 2011, “ Preliminary Mechanical Characterization of the Small Bowel for In Vivo Robotic Mobility,” ASME J. Biomech. Eng., 133(9), p. 091010. [CrossRef]
Samsom, M. , Smout, A. J. , Hebbard, G. , Fraser, R. , Omari, T. , Horowitz, M. , and Dent, J. , 1998, “ A Novel Portable Perfused Manometric System for Recording of Small Intestinal Motility,” Neurogastroenterol. Motil., 10(2), pp. 139–148. [CrossRef] [PubMed]
Connell, A. M. , 1961, “ The Motility of the Small Intestine,” Postgrad. Med. J., 37(434), pp. 703–716. [CrossRef] [PubMed]
Lee, Y. Y. , Erdogan, A. , and Rao, S. S. C. , 2014, “ How to Assess Regional and Whole Gut Transit Time With Wireless Motility Capsule,” J. Neurogastroenterol. Motil., 20(2), pp. 265–270. [CrossRef] [PubMed]
Kim, Y. , Lee, G. , Park, S. , Kim, B. , Park, J.-O. , and Cho, J.-H. , 2005, “ Pressure Monitoring System in Gastro-Intestinal Tract,” IEEE International Conference on Robotics and Automation (ICRA), Barcelona, Spain, Apr. 18–22, pp. 1321–1326.
Wang, W.-X. , Yan, G.-Z. , Sun, F. , Jiang, P.-P. , Zhang, W.-Q. , and Zhang, G.-F. , 2005, “ A Non-Invasive Method for Gastrointestinal Parameter Monitoring,” World J. Gastroenterol.: WJG, 11(4), pp. 521–524. [CrossRef]
Hoeg, H. D. , Slatkin, A. B. , Burdick, J. W. , and Grundfest, W. S. , 2000, “ Biomechanical Modeling of the Small Intestine as Required for the Design and Operation of a Robotic Endoscope,” IEEE International Conference on Robotics and Automation (ICRA), San Francisco, CA, Apr. 24–28, pp. 1599–1606.
Wang, X. , Meng, M. Q.-H. , and Chan, Y. , 2005, “ Physiological Factors of the Small Intestine in Design of Active Capsule Endoscopy,” 27th Annual Conference on IEEE Engineering in Medicine and Biology (EMBS), Shanghai, China, Sept. 1–4, pp. 2942–2945.
Miftahof, R. , and Akhmadeev, N. , 2007, “ Dynamics of Intestinal Propulsion,” J. Theor. Biol., 246(2), pp. 377–393. [CrossRef] [PubMed]
Woo, S. H. A. , Mohy-Ud-Din, Z. , and Cho, J. H. , 2013, “ Telemetry Capsule for Measuring Contractile Motion in the Small Intestine,” Biomed. Microdevices, 15(1), pp. 63–72. [CrossRef] [PubMed]
Li, P. , and Terry, B. S. , 2014, “ Design of a Swallowable Microrobotic Capsule for Measuring Small Intestine Pressure,” ASME J. Med. Devices, 8(3), p. 030910. [CrossRef]
Natali, C. D. , Beccani, M. , Obstein, K. L. , and Valdastri, P. , 2014, “ A Wireless Platform for In Vivo Measurement of Resistance Properties of the Gastrointestinal Tract,” Physiol. Meas., 35(7), pp. 1197–1214. [CrossRef] [PubMed]
Emmanuel, A. , and Roy, A. , 2007, “ Small Intestine and Colon Motility,” Medicine, 35(5), pp. 272–276. [CrossRef]
Li, P. , Kothari, V. , and Terry, B. S. , 2015, “ Design and Preliminary Experimental Investigation of a Capsule for Measuring the Small Intestine Contraction Pressure,” IEEE Trans. Biomed. Eng., 62(11), pp. 2702–2708. [CrossRef] [PubMed]
Li, P. , Kothari, V. , and Terry, B. S. , 2015, “ Processing and Analysis of Small Intestine Pressure Signal Based on Empirical Mode Decomposition,” ASME J. Med. Devices, 9(2), p. 020925.
Maqbool, S. , Parkman, H. P. , and Friedenberg, F. K. , 2009, “ Wireless Capsule Motility: Comparison of the SmartPill® GI Monitoring System With Scintigraphy for Measuring Whole Gut Transit,” Dig. Dis. Sci., 54(10), pp. 2167–2174. [CrossRef] [PubMed]
Saad, R. J. , and Hasler, W. L. , 2011, “ A Technical Review and Clinical Assessment of the Wireless Motility Capsule,” Gastroenterol. Hepatol., 7(12), pp. 795–804.
Terry, B. S. , Francisco, M. M. , Schoen, J. A. , and Rentschler, M. E. , 2013, “ Sensor for Measuring the Contact Force From Human Myenteric Contractions for In Vivo Robotic Capsule Endoscope Mobility,” ASME J. Med. Devices, 7(3), p. 030911. [CrossRef]
Böhm, B. , Milsom, J. W. , and Fazio, V. W. , 1995, “ Postoperative Intestinal Motility Following Conventional and Laparoscopic Intestinal Surgery,” Arch. Surg., 130(4), pp. 415–419. [CrossRef] [PubMed]
Miller, G. H. , 1926, “ The Effects of General Anesthesia on the Muscular Activity of the Gastrointestinal Tract A Study of Ether, Chloroform, Ethylene and Nitrous-Oxide,” J. Pharm. Exp. Thera., 27(1), pp. 41–59.
Szalai, M. , Kánainé, V. , Kiss, G. , Regőczi, H. , Horváth, Z. , and Rácz, I. , 2012, “ Small Bowel Transit Time in Patients Undergoing Capsule Endoscopy: Analyzis of Determining Factors,” Z. Gastroenterol., 50(5), p. A73.
Valdastri, P. , Menciassi, A. , and Dario, P. , 2008, “ Transmission Power Requirements for Novel ZigBee Implants in the Gastrointestinal Tract,” IEEE Trans. Biomed. Eng., 55(6), pp. 1705–1710. [CrossRef] [PubMed]
Fass, J. , Bares, R. , Hermsdorf, V. , and Schumpelick, V. , 1995, “ Effects of Intravenous Ketamine on Gastrointestinal Motility in the Dog,” Intensive Care Med., 21(7), pp. 584–589. [CrossRef] [PubMed]
Chan, F. P. , Li, K. C. , Heiss, S. G. , and Razavi, M. K. , 1999, “ A Comprehensive Approach Using MR Imaging to Diagnose Acute Segmental Mesenteric Ischemia in a Porcine Model,” Am. J. Roentgenol., 173(3), pp. 523–529. [CrossRef]
Duperret, S. , Lhuillier, F. , Piriou, V. , Vivier, E. , Metton, O. , Branche, P. , Annat, G. , Bendjelid, K. , and Viale, J. P. , 2007, “ Increased Intra-Abdominal Pressure Affects Respiratory Variations in Arterial Pressure in Normovolaemic and Hypovolaemic Mechanically Ventilated Healthy Pigs,” Intensive Care Med., 33(1), pp. 163–171. [CrossRef] [PubMed]
Allen, J. , 2007, “ Photoplethysmography and Its Application in Clinical Physiological Measurement,” Physiol. Meas., 28(3), pp. R1–R39. [CrossRef] [PubMed]
Miftahof, R. , and Fedotov, E. , 2005, “ Intestinal Propulsion of a Solid Non-Deformable Bolus,” J. Theor. Biol., 235(1), pp. 57–70. [CrossRef] [PubMed]
Terry, B. S. , Schoen, J. A. , and Rentschler, M. E. , 2012, “ Measurements of the Contact Force From Myenteric Contractions on a Solid Bolus,” J. Rob. Surg., 7(1), pp. 53–57. [CrossRef]
Stathopoulos, E. , Schlageter, V. , Meyrat, B. , de Ribaupierre, Y. , and Kucera, P. , 2005, “ Magnetic Pill Tracking: A Novel Non-Invasive Tool for Investigation of Human Digestive Motility,” Neurogastroenterol. Motil., 17(1), pp. 148–154. [CrossRef] [PubMed]
Worsøe, J. , Fynne, L. , Gregersen, T. , Schlageter, V. , Christensen, L. A. , Dahlerup, J. F. , Rijkhoff, N. J. , Laurberg, S. , and Krogh, K. , 2011, “ Gastric Transit and Small Intestinal Transit Time and Motility Assessed by a Magnet Tracking System,” BMC Gastroenterol., 11(1), p. 145. [CrossRef] [PubMed]

Figures

Grahic Jump Location
Fig. 1

(a) Interaction between the capsule and the small intestine (Pc : contact pressure and Pi : intraluminal pressure), (b) 3D model of the designed capsule, (c) assembly of pressure sensor 2, and (d) prototype of the capsule

Grahic Jump Location
Fig. 2

(a) Architecture of the pressure measurement system, (b) MCU board, (c) top view of IMU board, (d) bottom view of IMU board, and (e) receiver

Grahic Jump Location
Fig. 3

(a) Static test result (Pm : measured pressure by P2 and Pr : actual pressure applied on P2), (b) dynamic test result (Pm : measured pressure by P2 and Pr : actual pressure applied on P2), (c) transmission validation setup, and (d) animal test setup

Grahic Jump Location
Fig. 4

(a) Gathered pressure signals in the animal test and (b) contact pressure

Grahic Jump Location
Fig. 5

(a) One section of Pc with respiration and heartbeat induced pressure, (b) one section of Pc with contraction, respiration, and heartbeat induced pressure, and (c) fitting result with contact pressure model (Phb: heartbeat induced pressure; Pr: respiration induced pressure; Pt: tissue contraction induced pressure; Pc,m: model fitted contact pressure; and Pc,e: experimental contact pressure)

Grahic Jump Location
Fig. 6

(a) FFT of P1 and (b) FFT of Pc

Grahic Jump Location
Fig. 7

(a) Flowchart of the signal processing method, (b) isolated contraction induced pressure signal, and (c) FFT of contraction induced pressure signal

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In