Research Papers

A Microstructurally Driven Model for Pulmonary Artery Tissue

[+] Author and Article Information
Philip H. Kao, Steven R. Lammers, Lian Tian

Department of Mechanical Engineering, University of Colorado, Boulder, CO 80309

Kendall Hunter

Department of Pediatric Cardiology, University of Colorado Health Sciences, Denver, CO 80002

Kurt R. Stenmark

Department of Pediatrics, Developmental Lung Biology Laboratory, University of Colorado Health Sciences, Denver, CO 80002

Robin Shandas

Department of Mechanical Engineering, University of Colorado, Boulder, CO 80309; Center for Bioengineering, University of Colorado, Anschutz Medical Campus, Aurora, CO 80015

H. Jerry Qi1

Department of Mechanical Engineering, University of Colorado, Boulder, CO 80309qih@colorado.edu

1

Corresponding author.

J Biomech Eng 133(5), 051002 (Apr 08, 2011) (12 pages) doi:10.1115/1.4002698 History: Received July 26, 2009; Revised August 23, 2010; Posted October 04, 2010; Published April 08, 2011; Online April 08, 2011

A new constitutive model for elastic, proximal pulmonary artery tissue is presented here, called the total crimped fiber model. This model is based on the material and microstructural properties of the two main, passive, load-bearing components of the artery wall, elastin, and collagen. Elastin matrix proteins are modeled with an orthotropic neo-Hookean material. High stretch behavior is governed by an orthotropic crimped fiber material modeled as a planar sinusoidal linear elastic beam, which represents collagen fiber deformations. Collagen-dependent artery orthotropy is defined by a structure tensor representing the effective orientation distribution of collagen fiber bundles. Therefore, every parameter of the total crimped fiber model is correlated with either a physiologic structure or geometry or is a mechanically measured material property of the composite tissue. Further, by incorporating elastin orthotropy, this model better represents the mechanics of arterial tissue deformation. These advancements result in a microstructural total crimped fiber model of pulmonary artery tissue mechanics, which demonstrates good quality of fit and flexibility for modeling varied mechanical behaviors encountered in disease states.

FIGURES IN THIS ARTICLE
<>
Copyright © 2011 by American Society of Mechanical Engineers
Your Session has timed out. Please sign back in to continue.

References

Figures

Grahic Jump Location
Figure 1

Schematic of strips taken from excised pulmonary arteries

Grahic Jump Location
Figure 2

Schematic of crimped fiber model, the solid line is the undeformed fiber configuration and the dotted line is the deformed fiber configuration

Grahic Jump Location
Figure 3

The behavior of a single fiber under uniaxial extension is plotted. The nominal fiber stress P is normalized by the Young’s modulus to give the reduced fiber stress. (a) As the bending stiffness via R/l0 is increased, the fiber behavior goes from a sharp engagement to a broad engagement. The shape of the fiber is held constant at θ¯0=54 deg. (b) As the shape via θ¯0 is increased, the fiber behavior goes from a sharp transition with some fully developed stiffness to a broad transition with lower fully developed stiffness. The stretch at which the transition occurs also increases with increasing θ¯0. The radius of gyration ratio is kept constant where R/l0=0.05.

Grahic Jump Location
Figure 4

The visualized ellipsoidal structure tensor: This is generated by observing how the structure tensor transforms a unit vector. A longer dimension indicates a higher concentration of fibers in that direction. Note here that a0 and g0 are not aligned with the global coordinate system, but in the model, they are fixed to the circumferential and longitudinal directions, respectively.

Grahic Jump Location
Figure 5

The data from four sets of uniaxial tests and their corresponding model fits: Squares and triangles denote circumferential and axial data, respectively, while solid and dashed lines denote circumferential and axial data fits. (a)–(d) Here, the model’s versatility in different situations of material behaviors is seen. (e) Fiber material stretch as a function of material stretch for the case of circumferential test in (b).

Grahic Jump Location
Figure 6

(a) A parametric study varying γ, with κ=0.8. It is seen that when κ=γ (shown in circles), the behavior is transversely isotropic; the uniaxial stress-stretch curves lay on top of one another. As γ is increased in relation to κ (shown in squares (γ=0.95) and triangles (γ=0.9)), the degree of anisotropy is increased. The circumferential directions are in solid lines and the longitudinal directions are in dotted lines. The parameters held constant are KA=8×10−4nm−1, E=10 GPa, θ¯0=45 deg, and R/l0=0.1. (b) Results for the crimped fiber model only, showing the effect of changing crimped fiber parameter θ¯0. As θ¯0 increases, it pushes the engagement strain of the collagen further out. It also decreases the stiffness as the contour length is increased. The parameters held constant are KA=8×10−4nm−1, E=10 GPa, R/l0=0.1, κ=0.90, and γ=0.95. (c) The effect of changing crimped fiber parameter R/l0: As the radius of gyration is changed, it causes the transition to broaden and become more gradual. The parameters held constant are KA=8×10−4nm−1, E=10 GPa, θ¯0=36 deg, κ=0.90, and γ=0.95.

Grahic Jump Location
Figure 7

The engagement stretch as a function of (a) R/l0 and (b) θ¯0. The parameters used to calculate these data are μ=μa=μg=5 kPa, KA=12×10−4nm−1, E=10 GPa, θ¯0=45 deg, κ=0.90, and γ=0.95. The dotted lines denote typical parameter values.

Grahic Jump Location
Figure 8

Fits from the planar biaxial data: (a) Uniaxial stress-stretch data from the planar biaxial tests are shown with corresponding model fits. (b) The stress-stretch data from the 100:25 (C:L) experiment is shown with corresponding model prediction using parameters from the uniaxial fits. Data and corresponding fit for (c) circumferential uniaxial and (d) 100:25 (C:L) tests.

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In