Biomechanical Evaluation of a Novel Lumbosacral Axial Fixation Device

[+] Author and Article Information
Eric H. Ledet

Division of Orthopaedic Surgery, Albany Medical College, Albany, NY 12208 and Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY 12180

Michael P. Tymeson, Allen L. Carl

Division of Orthopaedic Surgery, Albany Medical College, Albany, NY 12208

Simon Salerno

Division of Neurosurgery, Albany Medical College, Albany, NY 12208

Andrew Cragg

Department of Radiology, University of Minnesota, Minneapolis, MN 55424

J Biomech Eng 127(6), 929-933 (Jul 14, 2005) (5 pages) doi:10.1115/1.2049334 History: Received April 01, 2005; Revised July 14, 2005

Background: Interbody arthrodesis is employed in the lumbar spine to eliminate painful motion and achieve stability through bony fusion. Bone grafts, metal cages, composite spacers, and growth factors are available and can be placed through traditional open techniques or minimally invasively. Whether placed anteriorly, posteriorly, or laterally, insertion of these implants necessitates compromise of the anulus—an inherently destabilizing procedure. A new axial percutaneous approach to the lumbosacral spine has been described. Using this technique, vertical access to the lumbosacral spine is achieved percutaneously via the presacral space. An implant that can be placed across a motion segment without compromise to the anulus avoids surgical destabilization and may be advantageous for interbody arthrodesis. The purpose of this study was to evaluate the in vitro biomechanical performance of the axial fixation rod, an anulus sparing, centrally placed interbody fusion implant for motion segment stabilization. Method of Approach: Twenty-four bovine lumbar motion segments were mechanically tested using an unconstrained flexibility protocol in sagittal and lateral bending, and torsion. Motion segments were also tested in axial compression. Each specimen was tested in an intact state, then drilled (simulating a transaxial approach to the lumbosacral spine), then with one of two axial fixation rods placed in the spine for stabilization. The range of motion, bending stiffness, and axial compressive stiffness were determined for each test condition. Results were compared to those previously reported for femoral ring allografts, bone dowels, BAK and BAK Proximity cages, Ray TFC, Brantigan ALIF and TLIF implants, the InFix Device, Danek TIBFD, single and double Harms cages, and Kaneda, Isola, and University plating systems. Results: While axial drilling of specimens had little effect on stiffness and range of motion, specimens implanted with the axial fixation rod exhibited significant increases in stiffness and decreases in range of motion relative to intact state. When compared to existing anterior, posterior, and interbody instrumentation, lateral and sagittal bending stiffness of the axial fixation rod exceeded that of all other interbody devices, while stiffness in extension and axial compression were comparable to plate and rod constructs. Torsional stiffness was comparable to other interbody constructs and slightly lower than plate and rod constructs. Conclusions: For stabilization of the L5-S1 motion segment, axial placement of implants offers potential benefits relative to traditional exposures. The preliminary biomechanical data from this study indicate that the axial fixation rod compares favorably to other devices and may be suitable to reduce pathologic motion at L5-S1, thus promoting bony fusion.

FIGURES IN THIS ARTICLE
<>
Copyright © 2005 by American Society of Mechanical Engineers
Your Session has timed out. Please sign back in to continue.

References

Figures

Grahic Jump Location
Figure 1

The axial fixation rod is a cannulated, threaded, titanium rod with two distinct threaded sections—one superior, one inferior

Grahic Jump Location
Figure 2

(a) A calf motion segment with intact vertebral bodies (VB) and intervertebral disk (IVD) is shown. (b) The motion segment was potted in a potting cup (PC) in low melting temperature alloy and tested in its intact state. (c) The inferior vertebral body was unpotted. (d) The segment was then inverted and the axial channel (AC) was drilled and repotted. (e) After drilled-only testing, the inferior vertebral body was again unpotted and an axial fixation rod (AFR) was inserted (shown partially inserted). A T-handle wrench (T) was used to place the rod through the vertebral body and intervertebral disk into the adjacent vertebral body, which remained potted in the potting cup.

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In