Possible Sources of Discrepancy Between Sphygmomanometer Cuff Pressure and Blood Pressure Quantified in a Collapsible-Tube Analogue

[+] Author and Article Information
C. D. Bertram, K. S. A. Butcher

Centre for Biomedical Engineering, University of New South Wales, Kensington, Sydney, Australia 2033

J Biomech Eng 114(1), 68-77 (Feb 01, 1992) (10 pages) doi:10.1115/1.2895452 History: Received August 06, 1991; Revised September 30, 1991; Online March 17, 2008

Abstract

This paper examines the assumption that the audible events detected as Korotkov sounds in sphygmomanometry occur when blood pressure equals arm-cuff pressure. Several effects that contribute to discrepancy between these pressures are quantified using an idealised arm-and-cuff system consisting of a thick-walled collapsible tube subject to external compression along a central part of its length. The effects studied are (1) transverse pressure difference, resulting from tissues sustaining a part of the external compression through (a) circumferential bending stiffness and (b) longitudinal curvature of the tensed localised neck at the site of initial collapse, (2) longitudinal pressure difference between upstream pressure and pressure at the collapse point due to both (a) viscous and (b) inertial pressure drop. These effects are found to compensate partially for each other; the pressure within the vessel at the collapse point is less than the cuff pressure, but is also less than the blood pressure at the upstream end of the cuff. All four of the contributing terms increase proportionally to the flow-rate raised to a power greater than one, except the viscous pressure drop. Owing to a progressive shortening of the collapsed neck as flow-rate increases, the viscous term is almost independent of the flow-rate. The overall discrepancy displays less flow-rate dependency and is smaller than some of the terms which contribute to it. This means that considerable accuracy is needed if measurements of the effects are to be used to correct the raw data on cuff pressure at the time of Korotkov sound emission so as to obtain an improved estimate of the blood pressure.

Copyright © 1992 by The American Society of Mechanical Engineers
Your Session has timed out. Please sign back in to continue.

References

Figures

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In