0
RESEARCH PAPERS

Measurement of Wall Shear Stress Distal to a Tri-Leaflet Valve in a Rigid Model of the Aortic Arch With Branch Flows

[+] Author and Article Information
S. Nandy, J. M. Tarbell

Department of Chemical Engineering, The Pennsylvania State University, University Park, PA 16802

J Biomech Eng 110(3), 172-179 (Aug 01, 1988) (8 pages) doi:10.1115/1.3108427 History: Received November 02, 1987; Revised May 25, 1988; Online June 12, 2009

Abstract

Flush mounted hot film anemometer probes were used to measure wall shear stress magnitudes on the inside and outside walls of a rigid model of the human aortic arch. The effects of the presence of an Ionescu-Shiley tri-leaflet bioprosthetic heart valve at the entrance of the aortic arch and the side flows through arteries located in the mid-arch region on wall shear stress magnitudes were determined. It was found that the presence of the tri-leaflet valve leads to an elevation of wall shear stress (relative to the same flow without a valve) over the entire aortic arch region by as much as 50 percent. The valve influence extended to about 180 deg from the entrance to the aorta on the inside wall and even further on the outside wall based on extrapolation of available data. Peak wall shear stress magnitudes measured on the outside wall were in the range of 1.5—4.0 N/m2 (15—40 dynes/cm2 ) over the length of the aortic arch and took on their highest values in the mid-arch region. Inside wall values were of comparable magnitude. It was observed that the presence of the aortic valve and side flow from the top of the aortic arch reduced wall shear stress reversal in the arch region.

Copyright © 1988 by ASME
Your Session has timed out. Please sign back in to continue.

References

Figures

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In