In-Vitro Fluid Dynamic Characteristics of the Abiomed Trileaflet Heart Valve Prosthesis

[+] Author and Article Information
Yi-Ren Woo, F. P. Williams, A. P. Yoganathan

Bio Fluid Dynamics Laboratory, School of Chemical Engineering, Georgia Institute of Technology, Atlanta, Ga. 30332

J Biomech Eng 105(4), 338-345 (Nov 01, 1983) (8 pages) doi:10.1115/1.3138430 History: Received May 12, 1982; Revised June 28, 1983; Online June 15, 2009

Abstract

The need for better and longer lasting trileaflet valves has led to the design and development of the Abiomed polymeric trileaflet valve prosthesis. In-vitro fluid dynamic studies on sizes 25 and 21 mm valves in the aortic position indicate an overall improvement in performance compared to the Carpentier-Edwards and Ionescu-Shiley tissue valves in current clinical use. The pressure drop studies yielded effective orifice areas of 1.99 and 1.54 cm2 , and performance indices of 0.41 and 0.45 for the Nos. 25 and 21 valves, respectively. Leaflet photography studies indicated that the two valve sizes had maximum opening areas of 225 and 145 mm2 , respectively, at a normal resting cardiac output. Steady and pulsatile flow velocity measurements with a laser-Doppler anemometer (LDA) system indicate that the flow field downstream of the Abiomed valve is jetlike and turbulent. Maximum mean square axial velocity fluctuations of 55 and 83 cm/s, and turbulent shear stresses of 220 and 450 N/m2 were measured in the immediate vicinity of the nos. 25 and 21 valves, respectively. The Abiomed valves studied had been originally configured for use in valved conduits, and it is therefore our opinion that further improvements can be made to the valve and stent design, which would enhance its fluid dynamic performance.

Copyright © 1983 by ASME
Your Session has timed out. Please sign back in to continue.

References

Figures

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In